EC9841T - Trace NO/NO2/NOx analyzer
|
The EC9841T NOX trace analyzer is a specially configured NOX analyzer designed to measure low levels of NO, NO2 & NOx, with high accuracy. |
![]() |
The EC9841T NOX trace analyzer combines microprocessor control with chemiluminescence to provide accurate measurements of NO, NO2 and NOX in the ranges of 50 ppt to 2000 ppb with a detection limit of 0.5 ppb.
Product page at http://www.ecotech.com/gas-analyzers/nox-trace-analyzer
Features
- Optional internal 12VDC power supply enables the NOX trace analyzer to be operated from a battery or solar powered source for remote or mobile applications.
- The micro processor facilitates external remote control and provides up to 100 channels of instrument operating parameters
- Data accessed via RS232, USB interface or optional ethernet connector
Specifications - EC9841T NOX trace analyzer
|
Range: LDL: Noise: Temperature range: Sample flow rate: Weight: Dimensions: Power: |
Auto-ranging for 0-2000 ppb 50 ppt with Kalman filter active 25 ppt or 0.2% of reading with kalman filter active 5 to 40oC 0.640 SLPM (Std) 20.9 kg 43.2 x 17.8 x 64.8 (w x h x d) 99-132 VAC, 198-264 VAC 47-63 Hz, 200 Watts |
- Supports internal data logging capability with storage up to 175 days of 5 minute data stored in flash ROM

