EC9830T - Trace CO analyzer
|
The EC9830T trace analyzer is Ecotech's specially configured carbon monoxide analyzer designed to measure low levels of CO with high accuracy. |
![]() |
The EC9830T trace CO analyzer incorporates Gas Filter Correlation (G.F.C) technology with infra red photometric detection. The analyzers sensitivity has been increased with the number of passes increased from 20 to 28. Detection of sub-ambient levels of CO is made challenging due to interference by H2O and CO2. The use of GFC technology minimises this interference and enables the instrument to measure CO in ambient air in the range of 0-20,000ppb.
CO background (trace) measurements in ambient air are usually made in the range of 25-350 ppb (levels that represent the range of global CO mixing ratios in the unpolluted atmosphere).
Product page at http://www.ecotech.com/gas-analyzers/co-trace-analyzer
Features
- Up to 100 backgrounds (zero) can be stored for diagnostic purposes
- Calibration curves are used to ensure that a linear signal output is achieved over two wide concentration ranges: 0-5000ppb and 5000-20,000 ppb
- Data accessed via RS232, USB interface or optional ethernet connector
Specifications - EC9830T CO trace analyzer
|
Range: LDL: Noise: Temperature range: Sample flow rate: Weight: Dimensions: Power: |
Auto ranging 0-20 ppm 25 ppb with Kalman or 300 sec filter active 12 ppb RMS with Kalman or 300 sec filter active 20 to 30oC, may be operated 15 to 35oC 1.00 SLPM (Std.), 2.00 LPM optional 20.9 kg 43.2 x 17.8 x 64.8 cm (w x h x d) 99-132 VAC, 198-264 VAC 47-63 Hz |
-
-
Supports internal data logging capability with storage up to 175 days of 5 minute data stored in flash ROM